DW-86L500

Aucma, DW-86L500

Suitable for hospitals, blood donation and blood centers, blood banks, scientific and research institutions, bioengineering as well as in the electronics and chemical industry.

Main features

    • CFC- free technology
    • CFC free refrigerant – R 290
    • LCD display with touch screen
    • Temperature graph on display
    • Integrated memory for 10 years
      Prepared for GSM alarm
    • Password protection
    • Optional: Wireless temperature monitoring kit with SMS notification system

Additional information

Model

DW-86L500

Capacity Net/ Gross

485/500L

Temperature range

-40~-86℃

Ambient temperature

10~32℃

Interior Dimension

620*655*1200mm

Exterior dimension

915*1055*1940mm

Climate class

4

Insulation thickness (mm)

130mm

Insulation layer

PU

Power supply (V)

230V

Frequency (Hz)

50Hz

Noise level (dB)

60

Power consumption (kWh/24h)

18

Type of defrost

Mmanual

Equipped with PT100 or NTC probe

PT100

Net/gross weigh

298/330KG

Exterior cabinet material

galvanized steel powder coated

Interior cabinet material

electric zinc plate

Door frame material

cold rolled steel sheet

Internal doors

tak

Refrigerant

R404A/R508B/R290

Number of compressors

2

Lock with key

yes

Thermometer hole design

yes

Protective gloves

yes

LCD display with touch screen

yes

Ambient temperature display

yes

USB connection

yes

Battery backup for alarms (approx.72 hrs )

yes

Direct download of logged data on USB memory stick

yes

Password protection. User's password / admin password

yes

Temperature graph on display

yes

Integrated data logger

yes

RS485/RS232 interface

yes

Integrated memory for 10 years

yes

Prepared for GSM alarm module

yes

Visual / Acoustic Alarm

yes

High, low temperature alarm

yes

Sensor malfuction alarm

yes

Open door alarm

yes

Power failure alarm

yes

Low battery alarm

yes

Condenser dirty alarm

yes

Over pressure alarm

yes

Over ambient temperature alarm

yes